Testomat 2000 iron in water

It is a colorimetric analyzer that controls and monitors the amount of iron in the water
All the materials used for this analyzer have been developed for this purpose. The dosage carried out by a piston pump is precise and accurate. The measuring chamber is mounted inside the equipment and has only two rings.
This prevents the growth of bacteria and makes maintenance easy. The consumption of the indicator and water are reduced.
Optimal for those companies that do not have control and / or maintenance personnel.

  • titrometric analysis,
  • dosage by piston pump
  • display of the result in various units
  • two maximum values can be programmed with output function (maximum value 1 exceeded, maximum value 2 exceeded)
  • programmable analysis interval between 0 – 99 minutes
  • possibility of volumetric analysis
  • 6 relay outputs: hardness value above the first programmed value, hardness value above the second programmed value, washing, malfunction alarm, maintenance request and AUX to control a chiller
  • 4 inputs: stop analysis, start of analysis, IN 1 zeroing the water meter and IN 2 for a water meter
  • documentation of analysis errors
  • additional programmable program (for example: to operate a chiller)
  • programmable maintenance address
  • galvanic division between high and low voltage
  • including pressure regulator 0.1 – 8 bar

accessory: output 0 (4) – 20mA, 0 / 2-10V or RS232 output

Technical data:

Power Sypply: 230V 50-60Hz, 1 – 8 bar (no.ord. 200210),
115V 50-60Hz, 1 – 8 bar (no.ord. 200212),
24V 50-60Hz, 1 – 8 bar (no.ord. 200211)
Power consumption max.: 30 VA
Protection calss: IP 65
Temperature max.: water/air 10 – 40 °C
Pressure max.: 1 – 8,0 bar (without flow regolator: 0,1 – 3 bar)
Dimensions: large 380, high 480, deep 280mm
Weight: 10,5 kg

Technical data

Nr. ord.
Testomat 2000 Fe 230V 50-60Hz, 0,1 – 8 bar
115V 50-60Hz, 0,1 – 8 bar
24V 50-60Hz, 0,1 – 8 bar
200290
200296
200295
Accessories: output 0 (4) – 20mA
output RS 232
output 0/2 – 10 V
cleaning set
200300
200309
200308
200012

Reagents for Testomat 2000 Iron (both reagents are needed):

Tipo ppm Fe No.ord.
Fe2005A
Fe2005B
0,00 – 0,10 200750
200755