It is a colorimetric analyzer that controls and monitors the amount of iron in the water
All the materials used for this analyzer have been developed for this purpose. The dosage carried out by a piston pump is precise and accurate. The measuring chamber is mounted inside the equipment and has only two rings.
This prevents the growth of bacteria and makes maintenance easy. The consumption of the indicator and water are reduced.
Optimal for those companies that do not have control and / or maintenance personnel.
Testomat Fe analyser
Testomat 2000 iron in water
- titrometric analysis,
- dosage by piston pump
- display of the result in various units
- two maximum values can be programmed with output function (maximum value 1 exceeded, maximum value 2 exceeded)
- programmable analysis interval between 0 – 99 minutes
- possibility of volumetric analysis
- 6 relay outputs: hardness value above the first programmed value, hardness value above the second programmed value, washing, malfunction alarm, maintenance request and AUX to control a chiller
- 4 inputs: stop analysis, start of analysis, IN 1 zeroing the water meter and IN 2 for a water meter
- documentation of analysis errors
- additional programmable program (for example: to operate a chiller)
- programmable maintenance address
- galvanic division between high and low voltage
- including pressure regulator 0.1 – 8 bar
accessory: output 0 (4) – 20mA, 0 / 2-10V or RS232 output
Technical data:
Power Sypply: | 230V 50-60Hz, 1 – 8 bar (no.ord. 200210), 115V 50-60Hz, 1 – 8 bar (no.ord. 200212), 24V 50-60Hz, 1 – 8 bar (no.ord. 200211) |
Power consumption max.: | 30 VA |
Protection calss: | IP 65 |
Temperature max.: | water/air 10 – 40 °C |
Pressure max.: | 1 – 8,0 bar (without flow regolator: 0,1 – 3 bar) |
Dimensions: | large 380, high 480, deep 280mm |
Weight: | 10,5 kg |
Technical data
Nr. ord. | ||
Testomat 2000 Fe | 230V 50-60Hz, 0,1 – 8 bar 115V 50-60Hz, 0,1 – 8 bar 24V 50-60Hz, 0,1 – 8 bar |
200290 200296 200295 |
Accessories: | output 0 (4) – 20mA output RS 232 output 0/2 – 10 V cleaning set |
200300 200309 200308 200012 |
Reagents for Testomat 2000 Iron (both reagents are needed):
Tipo | ppm Fe | No.ord. |
Fe2005A Fe2005B |
0,00 – 0,10 | 200750 200755 |
Want to know more about the Testomat 2000 Iron analyzer?
Contact us for any information
Contact us